Drive DeVilbiss® Vacu-Aide® QSU 7314 Suction Unit

DeVilbiss® Vacu-Aide® QSU 7314 Suction Unit

 

For more than half a century, DeVilbiss Healthcare has provided unmatched safety and performance in the airway suction device category. With the Vacu-Aide QSU, DeVilbiss brings you the same reliable, portable operation in a significantly quieter unit.

With more than 50% reduction in sound, the Vacu-Aide QSU is the quietest high flow/high suction portable unit on the market, making it a smart choice for patients from neonate to adult.

 

Product Highlights

  • Vacuum adjustment allows for 50–550 mm Hg and a free flow of 27 liters/minute
  • Meets American Association for Respiratory Care (AARC) guidelines for neonatal, infant, child and adult homecare suctioning
  • Sealed lead acid battery for use up to one-hour operation
  • Can return a patient to a state of comfort quickly – Meets ISO 10079-1:2009 high vacuum, high flow classification
  • New, innovative container assembly features integrated bacteria filter — Container and filter design maintains appropriate flow rate to meet ISO guidelines
  • Comes complete with internal battery, carrying case, 6′ tubing, AC/DC adapters and power cord
  • New Integrated gauge and flow regulator – Easy-to-turn knob conveniently placed around gauge for easy use and control
  • Manufactured and tested to ISO10079-1:1999 Standard for Suction Equipment, achieving outstanding performance in the following tests:
    • Electromagnetic Compatibility Test – Vacu-Aide does not interfere with or is not susceptible to interference from other equipment such as ventilators
    • Drop Test and Shock Vibration Test – Vacu-Aide is far less likely to be damaged when dropped or shipped
    • Excessive Temperature Test – Vacu-Aide effectively works in a wide range of environmental temperatures

 

AARC Guidelines for appropriate subatmospheric pressures:

  • Neonates: 60 – 80 mm Hg
  • Infants: 80 – 100 mm Hg
  • Children: 100 – 120 mm Hg
  • Adults: 100 – 150 mm Hg

Negative pressures should not exceed 150 mmHg as higher pressures have been shown to cause trauma, hypoxemia and atelectasis.

Source: American Association for Respiratory Care. Nasotracheal suctioning 2004 revision & update. Respir Care 2004 Sep;49(9): 1080-4 (62 references)

 

Specification

  • Dimensions (H x W x D): 21.1 x 20.3 x 21.6 cm
  • Weight : 3 kg
  • Vacuum range : 50 – 550 mmHg
  • Operating sound level: 55dBA
  • Air Flow : 27 l/min
  • Collection container capacity: 800ml (disposable), 1200ml (reusable)
  • Internal battery run time: 60 minutes
  • Operating temperature range: 0°C – 40°C
  • Electrical requirements: 100-240V, 50-60 Hz
  • ISO classification: EN ISO 10079-1 : 2009

 